We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

BACKGROUND/AIMS: To assess the diagnostic performance of retinal nerve fibre layer (RNFL) data of optical coherence tomography (OCT) for detecting glaucoma. METHODS: Secondary analyses of a prospective, multicentre diagnostic study (Glaucoma Automated Tests Evaluation (GATE)) referred to hospital eye services in the UK were conducted. We included data from 899 of 966 participants referred to hospital eye services with suspected glaucoma or ocular hypertension. We used both eyes' data and logistic regression-based receiver operator characteristics analysis to build a set of models to measure the sensitivity and specificity of the average and inferior quadrant RNFL thickness data of OCT. The reference standard was expert clinician examination including automated perimetry. The main outcome measures were sensitivity at 0.95 specificity and specificity at 0.95 sensitivity and the corresponding RNFL thickness thresholds. We explored the possibility of accuracy improvement by adding measures of within-eye and between-eye variation, scan quality, intraocular pressure (IOP) and age. RESULTS: Glaucoma was diagnosed in at least one eye in 17% of participants. Areas under the curve were between 0.83 and 0.88. When specificity was fixed at 0.95, the sensitivity was between 0.38 and 0.55, and the highest values were reached with models including the inferior quadrant rather than the average RNFL thickness. Fixing sensitivity at 0.95, the specificity was between 0.36 and 0.58. The addition of age, refractive error, IOP or within-subject variation did not improve the accuracy. CONCLUSION: RNFL thickness data of OCT can be used as a diagnostic test, but accuracy estimates remain moderate even in exploratory multivariable modelling of aiming to improve accuracy.

Original publication

DOI

10.1136/bjophthalmol-2017-310642

Type

Journal article

Journal

Br j ophthalmol

Publication Date

05/2018

Volume

102

Pages

604 - 610

Keywords

diagnostic tests/investigation, glaucoma, imaging, Adult, Age Factors, Aged, Diagnostic Techniques, Ophthalmological, Diagnostic Tests, Routine, Glaucoma, Humans, Intraocular Pressure, Logistic Models, Male, Middle Aged, Nerve Fibers, Prospective Studies, Retinal Ganglion Cells, Sensitivity and Specificity, Tomography, Optical Coherence