We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

The magnetic technique using a magnetic tracer and handheld magnetometer was successfully evaluated in breast surgery. Residual tracer at the injection site can cause susceptibility artefacts on breast magnetic resonance imaging (MRI), therefore for lesion localization a solid marker would be preferable. Four magnetic markers were developed for localization to evaluate its clinical applicability. Methods: Comparison was made of the maximum magnetic counts and artefact-volume on MRI. Results: All markers were successfully detected, the spring marker showed the highest mean magnetic counts (627.8 ± 400.2, mean ± SD) with unequal variances (p <.001) and the butterfly marker showed the smallest mean artefact-volume (11.1 ± 2.3 cm3) with a significant difference between markers (p =.049). Conclusion: Localization using a magnetic marker is feasible and further evaluation is required within a clinical trial.

More information Original publication

DOI

10.1016/j.jmmm.2018.01.030

Type

Journal article

Publisher

Elsevier

Publication Date

2018-01-16T00:00:00+00:00

Volume

460

Pages

334 - 339

Total pages

5