Cookies on this website
We use cookies to ensure that we give you the best experience on our website. If you click 'Continue' we'll assume that you are happy to receive all cookies and you won't see this message again. Click 'Find out more' for information on how to change your cookie settings.

We examined the relationships between the serum levels of chromium and cobalt ions and the inclination angle of the acetabular component and the level of activity in 214 patients implanted with a metal-on-metal resurfacing hip replacement. Each patient had a single resurfacing and no other metal in their body. All serum measurements were performed at a minimum of one year after operation. The inclination of the acetabular component was considered to be steep if the abduction angle was greater than 55 degrees. There were significantly higher levels of metal ions in patients with steeply-inclined components (p = 0.002 for chromium, p = 0.003 for cobalt), but no correlation was found between the level of activity and the concentration of metal ions. A highly significant (p < 0.001) correlation with the arc of cover was found. Arcs of cover of less than 10 mm were correlated with a greater risk of high concentrations of serum metal ions. The arc of coverage was also related to the design of the component and to size as well as to the abduction angle of the acetabular component. Steeply-inclined acetabular components, with abduction angles greater than 55 degrees, combined with a small size of component are likely to give rise to higher serum levels of cobalt and chromium ions. This is probably due to a greater risk of edge-loading.

Original publication

DOI

10.1302/0301-620x.90b10.20533

Type

Journal article

Journal

The Journal of bone and joint surgery. British volume

Publication Date

10/2008

Volume

90

Pages

1291 - 1297

Addresses

University Hospital Brussels, Laarbeeklaan 101, 1090 Brussels, Belgium. dr.desmet@heup.be

Keywords

Acetabulum, Hip Joint, Humans, Prosthesis Failure, Chromium, Cobalt, Arthroplasty, Replacement, Hip, Prosthesis Design, Equipment Failure Analysis, Hip Prosthesis, Adolescent, Adult, Aged, Middle Aged, Female, Male