We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

We report the results of a new technique of proximal cement fixation in 41 patients having revision hip arthroplasty using the Oxford modular femoral component. This consists of two parts, a metaphised wedge and a stem. The metaphyseal wedge is cemented and applies the load proximally to encourage bone healing. The wedge is free to slide on an uncemented stem that gains fixation distal to areas of bone loss, with the intention of reducing the risk of fracture. The clinical and radiographic follow-up ranged from a minimum of 3 years to 7 years. There have been no re-revisions. Ninety percent of the patients had pain or slight pain. Despite a relatively high rate of early subsidence, we feel that our results are encouraging, and justify the continued use of this component in hip revision surgery.

More information Original publication

DOI

10.1177/112070009800800201

Type

Journal article

Publisher

SAGE Publications

Publication Date

1998-01-01T00:00:00+00:00

Volume

8

Pages

62 - 69

Total pages

7