We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

OBJECTIVE: To determine the extent to which prostheses with no readily available evidence to support their use are being implanted in primary total hip arthroplasty. DESIGN: Systematic review of the literature. DATA SOURCES: The 9th annual report of the National Joint Registry of England and Wales (NJR) was analysed to identify prostheses with an Orthopaedic Data Evaluation Panel rating of "unclassified" or "pre-entry" used in primary total hip arthroplasty in 2011. A systematic review of those prostheses was carried out using PubMed, Cochrane, Embase, OVID, and Google databases. STUDY SELECTION: Prostheses used in primary total hip arthroplasty as published in the NJR's 9th annual report were analysed. Only literature that included the name of the prosthesis was included. Literature yielded in the search results was excluded if it reported animal, non-orthopaedic, non-total hip arthroplasty, or non-device related studies. RESULTS: The systematic review found that 24% (57/235) of all hip replacement implants available to surgeons in the UK have no evidence for their clinical effectiveness. It also shows that 10,617 (7.8%) of the 136,593 components used in primary hip replacements in 2011 were implanted without readily identifiable evidence of clinical effectiveness. These comprised 157 cemented stems (0.5% of 34,655 implanted), 936 (2.8% of 33,367) uncemented stems, 1732 (7.1% of 24,349) cemented cups, and 7577 (17.1% of 44,222) uncemented cups. CONCLUSIONS: This study shows that a considerable proportion of prostheses available to orthopaedic surgeons have no readily available evidence of clinical effectiveness to support their use. Concern exists about the current system of device regulation, and the need for a revised process for introducing new orthopaedic devices is highlighted.

Original publication

DOI

10.1136/bmj.f6956

Type

Journal article

Journal

Bmj

Publication Date

19/12/2013

Volume

347

Keywords

Arthroplasty, Replacement, Hip, Bone Cements, England, Evidence-Based Medicine, Hip Fractures, Hip Prosthesis, Prosthesis Design, Prosthesis Failure, Risk Factors, Treatment Outcome, Wales