We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

We previously showed that it is possible to cross-calibrate peripheral bone densitometers using the European Spine Phantom (ESP). We have now performed a multinational study of cross-calibrated radius bone density based on normal subjects of both sexes in eight European centers. Six centers were equipped with machines made by Scanco or Stratec for determining distal radial trabecular bone density by quantitative computed tomography (QCT) and two were equipped with Lunar SP2 single photon absorptiometry (SPA) equipment for measuring midshaft cortical bone density. Subjects recruited ranged from 20 to over 80 years of age. Over one hundred and fifteen men were studied by QCT and a different cohort of 104 men were studied with SPA; the equivalent figures for women were 235 and 123. Reference ranges were derived for bone density against age for each of the four groups, and their applicability is discussed in relation to between-center differences in the results obtained. There were insignificant differences (P > 0.05 with Bonferroni correction) between centers in the values obtained by QCT in the different populations. However, there were considerably larger and highly statistically significant differences between midshaft cortical bone density values of about 10% of overall means between subjects from eastern Finland and central Belgium (P < 0.001), with higher Finnish values. Women had considerably lower radial trabecular bone density values than men at all ages, a result that differentiates the radius from the spine. This sex difference widened after menopause. These results have important implications for understanding the contribution of bone density to the differential risk of Colles' fracture in the two sexes and suggest that further work is needed to establish young normal reference ranges for radial bone density in Europe.

Original publication

DOI

10.1007/BF02526878

Type

Journal article

Journal

Calcif tissue int

Publication Date

03/1996

Volume

58

Pages

135 - 143

Keywords

Absorptiometry, Photon, Adult, Aged, Aged, 80 and over, Bone Density, Bone and Bones, Calibration, Equipment and Supplies, Europe, Female, Humans, Male, Middle Aged, Models, Anatomic, Phantoms, Imaging, Radius, Reference Values, Software, Tomography, X-Ray Computed