We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

The aim of this work was to investigate the potential of a novel forward-viewing intravascular ultrasound (IVUS) system for flow quantification and colour flow imaging combined with B-mode imaging. A stiff 3.8-mm diameter catheter was used to scan a 72 degrees sector ahead of its tip. Operating at 30 MHz, the catheter was integrated with an IVUS scanner and a radiofrequency (RF) data-acquisition system. RF data were software processed for producing B-mode images and deriving velocity estimates. Steady flow in the range of 45 to 146 mL/min toward the catheter, was used in wall-less tissue-mimicking phantoms simulating healthy lumen (8-mm diameter), 30% diameter symmetrical stenosis and 37% diameter eccentric stenosis. The system provided colour flow images and good estimation of peak velocity and volumetric flows (within 1% to 9% and 16% to 48%, respectively, of calculated values) at 5 to 7 mm distal to the catheter. A sector forward-viewing IVUS imaging/Doppler system is suitable for combined anatomical and functional assessment of stenosed vessels.

Original publication

DOI

10.1016/s0301-5629(02)00699-3

Type

Journal article

Journal

Ultrasound med biol

Publication Date

01/2003

Volume

29

Pages

53 - 64

Keywords

Blood Flow Velocity, Blood Volume, Humans, Phantoms, Imaging, Regional Blood Flow, Ultrasonography, Doppler, Color, Ultrasonography, Interventional