We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

BACKGROUND: The limitations of serum testosterone and estradiol (E(2)) measurements using non-extraction platform immunoassays (IAs) are widely recognized. Switching to more specific mass spectrometry (MS)-based methods has been advocated, but directly comparative data on the two methods are scarce. METHODS: We compared serum testosterone and E(2) measurements in a large sample of middle-aged/elderly men using a common platform IA and a gas chromatography (GC)-MS method, in order to assess their limitations and advantages, and to diagnose male hypogonadism. Of subjects from the European Male Aging Study (n=3174; age 40-79 years), peripheral serum testosterone and E(2) were analyzed using established commercial platform IAs (Roche Diagnostics E170) and in-house GC-MS methods. RESULTS: Over a broad concentration range, serum testosterone concentration measured by IA and MS showed high correlation (R=0.93, P<0.001), which was less robust in the hypogonadal range (<11 nmol/l; R=0.72, P<0.001). The IA/MS correlation was weaker in E(2) measurements (R=0.32, P<0.001, at E(2) <40.8 pmol/l, and R=0.74, P<0.001, at E(2) >40.8 pmol/l). Using MS as the comparator method, IA ascertained low testosterone compatible with hypogonadism (<11 nmol/l), with 75% sensitivity and 96.3% specificity. The same parameters with IA for the detection of low E(2) (<40.7 pmol/l) were 13.3 and 99.3%, and for high E(2) (>120 pmol/l) 88.4 and 88.6%. CONCLUSION: A validated platform IA is sufficient to detect subnormal testosterone concentrations in the diagnosis of male hypogonadism. The IA used for E(2) measurements showed poor correlation with MS and may only be suitable for the detection of high E(2) in men.

Original publication

DOI

10.1530/EJE-11-1051

Type

Journal article

Journal

Eur j endocrinol

Publication Date

06/2012

Volume

166

Pages

983 - 991

Keywords

Adult, Aged, Aging, Cohort Studies, Electrochemistry, Estradiol, Europe, Gas Chromatography-Mass Spectrometry, Humans, Hypogonadism, Immunoassay, Luminescent Measurements, Male, Middle Aged, Sensitivity and Specificity, Serum Albumin, Sex Hormone-Binding Globulin, Testosterone